86-0536-8882360
X-Ray High Voltage Cables
Product Highlights
√ For medical applications, such as x-ray machine, DR, CT etc.
√ From basic radiography systems to high performance systems.
√ Customized for highest flexibility.
√ Complete product range.
√ Every cable length available.
Configuration
Cable Length
Standard 8M/12M
Customized 15M/20M
Technical Parameters
Electrical Data | |
Rated Voltage | 90kVdc |
Shield Resistance | 0.010 Ohm/m |
Coverage Shielding Braid | > 95% |
Cable Capacity | < 120pF/m |
Test Voltages | |
High Voltage Insulation | 90kVdc |
Low Voltage Insulation | 500 Vac |
Mechanical Data: Cable | |
Outer Diameter | 16.5 mm + 0.5 |
Bending Radius | > 150mm |
Low Voltage Conductor | 2×1.5mm2 |
Common Conductor | 2×0.75mm2 |
Conductor Resistance | 0.013 Ohm/m |
Cable Sheat Colour | Light grey |
Mechanical Data: Plug | |
Multiple Contact Pin | 8 point contact |
Max. Temperature | 110 ℃ |
Flanges | Demountable |
Extremely Short Sleeve | 34 mm |
Small Plug Diameter | 40 mm |
Solderless Connection | (EMC grounding) |
Detail Pictures
To prevent excessive bending. Its bending radius should not less than the 5 to 8 times of the cable diameter so as not to cause cracks and reduce the insulation strength. Please keep the cables dry and clean in normal time, and avoid oil, moisture and the harmful gas erosion so as to avoid the aging of the rubber.
You May Be Interested In:L01 X-Ray Exposure Hand Switch