86-0536-8882360
X-Ray High Voltage Cables
Configuration
Cable Length | ||
Standard | 8m | 12m |
Customized | 15m | 20m |
Specification
Electrical Data | |
Rated Voltage | 90kVdc |
Shield Resistance | 0.010 Ohm/m |
Coverage Shielding Braid | > 95% |
Cable Capacity | < 120pF/m |
Test Voltages | |
High Voltage Insulation | 90kVdc |
Low Voltage Insulation | 500 Vac |
Mechanical Data: Cable | |
Outer Diameter | 16.5 mm + 0.5 |
Bending Radius | > 150mm |
Low Voltage Conductor | 2×1.5mm2 |
Common Conductor | 2×0.75mm2 |
Conductor Resistance | 0.013 Ohm/m |
Cable Sheat Colour | Light grey |
Mechanical Data: Plug | |
Multiple Contact Pin | 8 point contact |
Max. Temperature | 110 ℃ |
Flanges | Demountable |
Extremely Short Sleeve | 34 mm |
Small Plug Diameter | 40 mm |
Solderless Connection | (EMC grounding) |
FAQ -- Technical Knowledge of High Voltage Cables
Q: What is the working principle of high voltage cables for x-ray machine?
A: The high voltage cables are mainly used on large and midium scale x-ray machine, which is connected with the high voltage generator and x-ray tube to transfer the high voltage from generator to the anode and cathode of the x-ray tube, and transfer the heating voltage to the x-ray tube filament.
Q: What is the operation attention of high voltage cables
A: To prevent excessive bending. Its bending radius should not less than the 5 to 8 times of the cable diameter so as not to cause cracks and reduce the insulation strength. Please keep the cables dry and clean in normal time, and avoid oil, moisture and the harmful gas erosion so as to avoid the aging of the rubber.
You may be interested in:Weifang Newheek Electronic Technology Co., Ltd.